Synchronisation d'un interféromètre Fabry-Perot pour l'infrarouge (5,5μ à 8,5μ) et d'un spectromètre à réseau
D. Laplaze et A.M. VergnouxDépartement de Physique Cristalline. Faculté des Sciences, Montpellier
Abstract
In order to improve the performance of a grating spectrometer, we have added a Fabry-Perrot interferometer used from 5,5 μ to 8,5 μ. We have built a mechanical system to synchronize both the sweeping of the interferometer and the wavelength-pass of the grating spectrometer, for any resolving power, chosen in a wide region (the interference order can varied in the ratio from 1 to 6). We have analysed with this arotational-vibrational band of water-vapour about 7.5 μ, and we have oblained a gain between 4 and 5 for the product L.R.
Résumé
Afin d'améliorer les performances d'un spectromètre à réseau, nous lui avons adjoint un interféromètre Fabry-Perot, fonctionnant entre 5,5 μ et 8,5 μ. Nous avons construit un entrainement mécanique permettant de synchroniser le balayage de l'interféromètre et le défilement des longueurs d'onde du prémonochromateur quelle que soit la résolvance choisie dans un large domaine. (L'ordre d'interférence pouvant varier dans le rapport de 1 à 6). Nous avons analysé, avec cette chaine, une bande de vibration-rotation de l'eau vers 7,5 μ et nous avons obtenu un gain de 4 à 5 sur le produit L. R.
0760R - Visible and ultraviolet spectrometers.
Key words
light interferometers -- spectrometers